Quality Control Testing of Dose Calibrators

R. Y. Chu,W. Simon
1996-06-01
Abstract:We have reviewed quality control testing of dose calibrators and have analyzed the potential for error. The uncertainty in half-lives of reference sources can lead to inaccurate assessment of the accuracy and constancy in dose calibrator performance. Molybdenum-99 contamination of '"Tc used as a reference source can appear as a nonlinear response of the dose calibrator. The useful life of a reference source and the tolerable amount of Mo contamination are examined. Key Wonts: dose calibrator; accuracy; constancy; linearity; geometry; quality control