HPLC method for simultaneous analysis of ticagrelor and its organic impurities and identification of two major photodegradation products

Lívia Maronesi Bueno,Joanna Wittckind Manoel,Camila Ferrazza Alves Giordani,Andreas Sebastian Loureiro Mendez,Nadia Maria Volpato,Elfrides Eva Scherman Schapoval,Martin Steppe,Cássia Virginia Garcia
DOI: https://doi.org/10.1016/j.ejps.2016.11.004
2017-01-15
Abstract:A simple, fast and sensitive analytical method by high-performance liquid chromatography (HPLC) was developed and validated for the simultaneous determination of ticagrelor and two synthesis impurities. The HPLC method was established using an Agilent 1200 Series equipment coupled to photodiode array detector (PDA) at 270nm with a Zorbax Plus C8 column (150×4.6mm, 5.0μm), injection volume of 20μL, and a constant temperature of 25°C. The mobile phase consisted of acetonitrile: ammonium acetate 50mM (57:43, v/v) and pH adjusted to 8.2 with ammonium hydroxide 6M, at a flow rate of 0.7mL/min. No interference peaks from excipients and diluent system indicated the specificity of the method. The calibration curves showed determination coefficients (r2)>0.99, calculated by linear regression. The limit of quantitation (LOQ) for impurities 1 and 2 were 2.0 and 0.2μg/mL, respectively. Intra and interday relative standard deviations (RSDs) were <2% for ticagrelor and <6% for the impurities, proving the precision of the method. Besides, two mayor degradation products formed when sample solutions of ticagrelor were exposed to UVC radiation were elucidated and the mechanisms involved in the photolytic degradation of ticagrelor were proposed.