An unusual cause of increasing airway pressure during anesthesia

L V Klein,D V Wilson
DOI: https://doi.org/10.1111/j.1532-950x.1989.tb01078.x
Abstract:A mare with a thoracic laceration was anesthetized with halothane-O2 for surgical repair. Shortly after initiation of controlled ventilation, inspiratory plateau pressure began to increase. The increase in pressure continued despite decreasing the tidal volume. Tension pneumothorax was suspected, but because arterial PCO2 and PO2 were normal, the anesthetic apparatus was examined. Excess gas continued to exit through the bellows pop-off valve even when the fresh gas flow was turned off. It was determined that driving gas from the ventilator was entering the breathing circuit through a hole in the bellows. Testing the integrity of the bellows before use is suggested to avoid this potentially harmful complication.