Rapid patient-specific FEM meshes from 3D smart-phone based scans

Ethan K Murphy,Joel Smith,Michael A Kokko,Seward B Rutkove,Ryan J Halter
DOI: https://doi.org/10.1088/1361-6579/ad26d2
2024-02-08
Physiological Measurement
Abstract:Objective: The objective of this study was to describe and evaluate a smart-phone based method to rapidly generate subject-specific finite element method (FEM) meshes. More accurate FEM meshes should lead to more accurate thoracic electrical impedance tomography (EIT) images. Approach: The method was evaluated on an iPhone® that utilized an app called Heges, to obtain 3D scans (colored, surface triangulations), a custom belt, and custom open-source software developed to produce the subject-specific meshes. The approach was quantitatively validated via mannequin and volunteer tests using an infrared tracker as the gold standard, and qualitatively assessed in a series of tidal-breathing EIT images recorded from 9 subjects. Main Results: The subject-specific meshes can be generated in as little as 6.3 minutes, which requires on average 3.4 minutes of user interaction. The mannequin tests yielded high levels of precision and accuracy at3.2±0.4 mm and 4.0±0.3 mm root mean square error (RMSE), respectively. Errors on volunteers were larger (12.1±7.0 mm RMSE precision and 14.7±5.4 mm RMSE accuracy), but could be improved with additional data collection strategies. Significance: Easy-to-generate, subject-specific meshes could be utilized in the thoracic EIT community, potentially reducing geometric-based artifacts and improving the clinical utility of EIT.
engineering, biomedical,biophysics,physiology