Maximizing NMR Sensitivity: A Guide to Receiver Gain Adjustment

Josh P. Peters,Jan-Bernd Hövener,Andrey N. Pravdivtsev
2024-09-26
Abstract:Nuclear magnetic resonance is revolutionizing in parallel with the development of methodology and all electronics and hardware around it. The primary objective of such modifications is to make the system more reliable and sensitive. Almost a decade ago, Bruker introduced an Avance NEO console with an extended dynamic range of the receiver gain (RG). The system offered a maximum sensitivity even at very low RG. This feature is handy in the case of hyperpolarization when too much signal can cause clipping of the free induction decay signal. We analyzed the RG profile for our spectrometer's 1H, 2H, 13C, and 15N nuclei and found nonmonotonous sensitivity behavior characterized by signal-to-noise ratio (SNR). For example, 13C SNR at RG = 18 is 5.5% higher than SNR at maximum RG = 101 and 51.6% higher than at RG = 20.2. Having found this, we advertise here that other NMR users with any spectrometers should calibrate their SNR vs. RG and find the RG value right for their experiments, as automatic RG adjustment does not know about the actual performance of each system. We offer a tool and strategy for such an optimization and exemplify the benefits of using SNR-adjusted RG value with hyperpolarized [1-13C]pyruvate measurements.
Instrumentation and Detectors,Chemical Physics