Accurate, precise pressure sensing with tethered optomechanics

Olivia R. Green,Yiliang Bao,John R. Lawall,Jason J. Gorman,Daniel S. Barker
2024-08-31
Abstract:We show that optomechanical systems can be primary pressure sensors with uncertainty as low as 1.1 % of reading via comparison with a pressure transfer standard. Our silicon nitride and silicon carbide sensors are short-term and long-term stable, displaying Allan deviations compatible with better than 1 % precision and baseline drift significantly lower than the transfer standard. We also investigate the performance of optomechanical devices as calibrated gauges, finding that they can achieve total uncertainty less than 1 %. The calibration procedure also yields the thin-film density of our sensors with state-of-the-art precision, aiding development of other calibration-free optomechanical sensors. Our results demonstrate that optomechanical pressure sensors can achieve accuracy, precision, and drift sufficient to replace high performance legacy gauges.
Applied Physics,Instrumentation and Detectors
What problem does this paper attempt to address?