Batch VUV4 Characterization for the SBC-LAr10 scintillating bubble chamber

H. Hawley-Herrera,E. Alfonso-Pita,E. Behnke,M. Bressler,B. Broerman,K. Clark,J. Corbett,C.E. Dahl,K. Dering,A. de St. Croix,D. Durnford,P. Giampa,J. Hall,O. Harris,N. Lamb,M. Laurin,I. Levine,W.H. Lippincott,X. Liu,N. Moss,R. Neilson,M.-C. Piro,D. Pyda,Z. Sheng,G. Sweeney,E. Vázquez-Jáuregui,S. Westerdale,T.J. Whitis,A. Wright,E. Wyman,R. Zhang
2024-07-23
Abstract:The Scintillating Bubble Chamber (SBC) collaboration purchased 32 Hamamatsu VUV4 silicon photomultipliers (SiPMs) for use in SBC-LAr10, a bubble chamber containing 10~kg of liquid argon. A dark-count characterization technique, which avoids the use of a single-photon source, was used at two temperatures to measure the VUV4 SiPMs breakdown voltage ($V_{\text{BD}}$), the SiPM gain ($g_{\text{SiPM}}$), the rate of change of $g_{\text{SiPM}}$ with respect to voltage ($m$), the dark count rate (DCR), and the probability of a correlated avalanche (P$_{\text{CA}}$) as well as the temperature coefficients of these parameters. A Peltier-based chilled vacuum chamber was developed at Queen's University to cool down the Quads to $233.15\pm0.2$~K and $255.15\pm0.2$~K with average stability of $\pm20$~mK. An analysis framework was developed to estimate $V_{\text{BD}}$ to tens of mV precision and DCR close to Poissonian error. The temperature dependence of $V_{\text{BD}}$ was found to be $56\pm2$~mV~K$^{-1}$, and $m$ on average across all Quads was found to be $(459\pm3(\rm{stat.})\pm23(\rm{sys.}))\times 10^{3}~e^-$~PE$^{-1}$~V$^{-1}$. The average DCR temperature coefficient was estimated to be $0.099\pm0.008$~K$^{-1}$ corresponding to a reduction factor of 7 for every 20~K drop in temperature. The average temperature dependence of P$_{\text{CA}}$ was estimated to be $4000\pm1000$~ppm~K$^{-1}$. P$_{\text{CA}}$ estimated from the average across all SiPMs is a better estimator than the P$_{\text{CA}}$ calculated from individual SiPMs, for all of the other parameters, the opposite is true. All the estimated parameters were measured to the precision required for SBC-LAr10, and the Quads will be used in conditions to optimize the signal-to-noise ratio.
Instrumentation and Detectors
What problem does this paper attempt to address?