Design and commissioning of a high-level control system for a medical isochronous cyclotron

Petter Hofverberg,Jean-Marc Bergerot,Jean-Michel Bruneau,Richard Trimaud
DOI: https://doi.org/10.18429/JACoW-IPAC2024-THPG15
2024-09-26
Abstract:MEDICYC (MEDical CYClotron) is an isochronous cyclotron dedicated to radiotherapy which was built and commissioned in Nice, France, in 1990 by a local team aided by experts from CERN. The cyclotron accelerates negative H to a maximum energy of 65 MeV and uses stripping to extract a proton beam. Its primary purpose is treating ocular melanoma by protontherapy but a significant research activity is also present on beam-lines dedicated for this purpose. An extensive refurbishment program of the cyclotron has been started to cope with the end-of-life and/or the obsolescence of several sub-systems. In this context, a new high-level cyclotron control system has been developed and commissioned in 2021-2024. The primary responsibility of the system is the high-level coordination of the source, the RF system, the beam-line and cyclotron magnets, to produce and deliver a beam with a given set of characteristics. A secondary responsibility is the collection, visualization and analysis of sub-system and beam data for monitoring and pre-emptive fault detection. In this contribution, the control system software architecture is presented and the infrastructure on which the systems are deployed is laid out.
Accelerator Physics,Instrumentation and Detectors
What problem does this paper attempt to address?
The paper aims to address the aging issues of the MEDICYC (Medical Cyclotron) control system, particularly in the context of a large-scale overhaul required due to system components nearing the end of their life or becoming obsolete. The original control system was a large monolithic application that was difficult to understand and maintain, and not easily adaptable to the evolution of subsystems. Therefore, the main objective of the paper is to design and implement a new advanced control system to tackle these challenges. The main goals of the new system include: 1. **Minimize impact on the advanced control system**: When subsystems or hardware components change, the impact on the entire system should be minimized. 2. **Support degraded mode operation**: The cyclotron should still be able to operate when certain hardware components are temporarily unavailable, reconfigured, or swapped. 3. **Provide extensive data logging and analysis tools**: For system monitoring and preventive measures, improving the system's reliability and stability. 4. **Support multiple beamlines and beam characteristics**: To meet different research and treatment needs. By achieving these goals, the new control system not only enhances the system's flexibility and reliability but also provides better support for future hardware upgrades and system maintenance.