New infrared camera of the Caucasian Mountain Observatory of the SAI MSU: design, main parameters, and first light

S.G. Zheltoukhov,A.M. Tatarnikov,A.A. Belyakova,E.A. Koksharova
2024-04-02
Abstract:This paper presents a prototype of an infrared photometer, created at SAI of MSU based on the commercial infrared module Gavin-615A. The operating spectral range of the photometer is 3-5 mkm. Investigations of the photometer's detector have shown that its parameters coincide with those stated by the manufacturer. The nonlinearity of the detector does not exceed approx 5% across the entire signal range, and coefficients for correction functions were determined. Additionally, we determined the readout noise RN=1200+-210 e-, the conversion coefficient GAIN=520+-9 e-/ADU, the signal magnitude of the bias frame BIAS = 960.5+-2.2 ADU, and the dark current 9.3e6 +-1.1e6 e-/s, which is the sum of the detector's dark current and the radiation from the entrance window of the detector module. The value of dark current was measured at a window temperature of 6C. Observations commenced with the photometer at the 2.5-m telescope of the Caucasian Mountain Observatory of MSU, with the first results presented in this paper. The unvignetted field of view was 30". In the M band under good atmospheric conditions, an image quality close to the diffraction limit was achieved. Images of a star with brightness L=7.96 and M=6.78 were obtained over the 20-second exposure time and a SNR=10. It is shown that at high image quality with SNR=3 and exposure of 20 seconds it is possible to observe stars up to L=9 mag and M=8 mag. The main module of the photometer was also used in measurements of the sky background brightness.
Instrumentation and Methods for Astrophysics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to develop and test a new type of infrared camera in order to expand the observational capabilities of the 2.5 - meter telescope at the Caucasus Mountain Observatory in the 3 - 5 micrometer band. Specifically, the research aims to: 1. **Design and Construction**: Design and build a prototype infrared photometer (LMP) based on the commercial infrared module Gavin - 615A for astronomical observations. 2. **Performance Evaluation**: Measure and evaluate in detail the key parameters of the LMP photometer, including nonlinearity, gain coefficient (GAIN), read - out noise (RN), bias frame (BIAS), and dark current, etc. 3. **First Observation**: Use the newly - built LMP photometer for the first observation, verify its performance, and demonstrate its ability to observe celestial bodies in different bands (such as L and M bands). 4. **Background Brightness Measurement**: Use the LMP photometer to measure the sky background brightness and provide reference data for future observations. ### Key Issues 1. **Expanding the Observation Band**: Modern astronomical research trends and the climate conditions of observatories allow observations in the long - wave infrared region, which is very important for studying cold objects, dust clouds, and other celestial bodies. 2. **Improving the Observation Capability**: Improve the research capabilities of different types of celestial bodies (such as young stars, AGB and post - AGB stars, comets, bright active galactic nuclei, and celestial bodies with interstellar dust shells) by improving detector technology and optimizing instrument design. 3. **Reducing Background Noise**: Reduce the instrument background noise to improve the signal - to - noise ratio, so as to be able to observe fainter celestial bodies. 4. **Practical Application**: Ensure that the LMP photometer can work stably under real - world observation conditions and is compatible with existing telescope software and equipment. ### Main Results - **Performance Parameters**: - Non - linearity does not exceed 5%. - Gain coefficient \( GAIN = 520\pm9 \, e^-/\text{ADU} \) - Read - out noise \( RN = 1200\pm310 \, e^- \) - Bias frame \( BIAS = 960.5\pm2.2 \, \text{ADU} \) - Dark current \( \approx (9.3\pm1.1)\times 10^6 \, e^-/\text{s} \) - **First Observation**: - Successfully observed celestial bodies such as the carbon star CW Leo, Jupiter, Mars, and Venus. - Under good atmospheric conditions, achieved an image quality close to the diffraction limit (FWHM ≈ 0.6 - 0.7"). - **Background Brightness Measurement**: - Built a special device for sky scanning and measured the sky background brightness at different elevation angles. In conclusion, this paper demonstrates the design, performance evaluation of the LMP photometer and its application potential in actual astronomical observations, providing important tools and data support for further infrared astronomical research.