Strategies and safety simulations for ultrasonic cervical spinal cord neuromodulation

Rui Xu,Sven Bestmann,Bradley E. Treeby,Eleanor Martin
2024-02-22
Abstract:Focused ultrasound spinal cord neuromodulation studies have demonstrated spinal cord neuromodulation in small animals. The safe and efficacious translation of these approaches to human scale requires an understanding of ultrasound propagation and heat deposition within the human spine. To address this, combined acoustic and thermal modelling was used to assess the pressure and heat distributions produced by a 500 kHz source focused to the C5/C6 level of the cervical spine via two approaches a) the posterior acoustic window between vertebral posterior arches, or b) the lateral intervertebral foramen from which the C6 spinal nerve exits. Pulse trains of 150 0.1 s pulses with a pulse repetition frequency of 0.33 Hz and free-field spatial peak pulse-averaged intensity of 10 W/cm^2 were simulated for the CT volumes of four subjects and for $\pm$10 mm translational and $\pm$10° rotational source positioning errors. Target pressures ranged between 20% and 70% of free-field spatial peak pressures with the posterior approach, and 20% and 100% with the lateral approach. When the source was optimally positioned with the posterior approach, peak spine heating values were below 1°C, but source mis-positioning resulted in bone heating up to 4°C. Heating with the lateral approach did not exceed 2°C within the mispositioning range. There were substantial inter-subject differences in target pressures and peak heating values. Target pressure varied three to four-fold between subjects, depending on approach, while peak heating varied approximately two-fold between subjects. This results in a near ten-fold range in the target pressure achieved per degree of peak heating between subjects. This highlights the importance of developing trans-spine ultrasound simulation software for the assurance of subject-specific safety and efficacy of focused ultrasound spinal cord therapies.
Medical Physics
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to evaluate the safety and effectiveness of low - intensity focused ultrasound in the neuromodulation of the human cervical spinal cord. Specifically, the study evaluates the effects of two different ultrasonic transmission paths (posterior and lateral paths) on the C5/C6 level of the cervical spine by combining acoustic and thermal simulations. The following is a detailed problem description: 1. **Safety evaluation**: - The study aims to ensure that before conducting human experiments, the propagation characteristics of ultrasound in the human spine and the heat deposition situation are understood. - Special attention is paid to the influence of the positioning error of the ultrasonic source on the heating of bone tissue. 2. **Effectiveness evaluation**: - Evaluate the pressure distribution generated by ultrasound in the target area (cervical spinal cord) under different paths (posterior and lateral paths). - Compare the effectiveness of different paths to determine which method is more suitable for clinical applications. 3. **Influence of individual differences**: - The study also explores the influence of anatomical structure differences between different subjects on the propagation and heating effects of ultrasound. - Analyze the variation ranges of the target pressure and peak heating value among different individuals. ### Specific problem description - **Ultrasound propagation and heat deposition**: The study used an ultrasonic source of 500 kHz and simulated the sound pressure and temperature distributions under different pulse sequence parameters (such as a pulse repetition frequency of 0.33 Hz and a free - field spatial - peak pulse - average intensity of 10 W/cm²). - **Path selection**: - **Posterior path**: Focus the ultrasound on the spinal cord at the C5/C6 level through the acoustic window between the vertebral arches. - **Lateral path**: Focus the ultrasound on the C6 spinal nerve root through the intervertebral foramen. - **Influence of positioning error**: Simulate the influence of translational errors of ±10 mm and rotational errors of ±10° on the propagation and heating effects of ultrasound. ### Key findings - **Safety**: - Under the posterior path, when the ultrasonic source is correctly positioned, the maximum temperature rise of the spinal cord does not exceed 1°C; however, the positioning error may cause the bone tissue to heat up to 4°C. - Under the lateral path, even if there is a positioning error, the maximum temperature rise does not exceed 2°C. - **Effectiveness**: - The target pressure range of the posterior path is 20% - 70% of the free - field spatial - peak pressure, while the target pressure range of the lateral path is 20% - 100%. - There are significant individual differences among different subjects, and the variation ranges of the target pressure and peak heating value are large. ### Conclusion This study emphasizes the importance of developing cross - spine ultrasound simulation software to ensure the safety and treatment effect of each subject. This provides an important reference basis for future human experiments.