Measurement of the Photon Detection Efficiency of Hamamatsu VUV4 SiPMs at Cryogenic Temperature

Rodrigo Álvarez-Garrote,Enrique Calvo,Aritz Canto,José Ignacio Crespo-Anadón,Clara Cuesta,Andrés de la Torre Rojo,Inés Gil-Botella,Sergio Manthey Corchado,Iván Martín,Carmen Palomares,Laura Pérez-Molina,Antonio Verdugo de Osa
DOI: https://doi.org/10.1016/j.nima.2024.169347
2024-04-02
Abstract:Liquid argon time projection chambers (TPC) are widely used in neutrino oscillation and dark matter experiments. Detection of scintillation light in liquid argon TPC's is challenging because of its short wavelength, in the VUV range, and the cryogenic temperatures (~86 K) at which the sensors must operate. Wavelength shifters (WLS) are typically needed to take advantage of the high Photon Detection Efficiency (PDE) in the visible range of most of photondetectors. The Hamamatsu VUV4 S13370--6075CN SiPMs can directly detect VUV light without the use of WLS, which main benefit is an improved PDE at these short wavelengths, but also the visible light from WLS. The manufacturer (Hamamatsu Photonics K.K.) provides a complete characterization of these devices at room temperature; however, previous studies have indicated a decrease of the PDE at cryogenic temperature for VUV light. In this work, we present the measurement of the PDE of VUV4 SiPMs at cryogenic temperature for different wavelengths in the range [270, 570] nm. A dedicated measurement at 127 nm is also shown.