Performance of PMTs for the JSNS2 experiment
J. S. Park,H. Furuta,T. Maruyama,S. Monjushiro,K. Nishikawa,M. Taira,J. S. Jang,K. K. Joo,J. Y. Kim,I. T. Lim,D. H. Moon,J. H. Seo,C. D. Shin,A. Zohaib,P. Gwak,M. Jang,S. Ajimura,T. Hiraiwa,T. Nakano,M. Nomachi,T. Shima,Y. Sugaya,M. K. Cheoun,J. H. Choi,M. Y. Pac,T. Dodo,Y. Hino,F. Suekane,R. Ujiie,M. Harada,S. Hasegawa,Y. Kasugai,S. Meigo,K. Sakai,S. Sakamoto,K. Suzuya,J. R. Jordan,J. Spitz,E. Marzec,M. Botran,T. Kawasaki,T. Konno,H. I. Jang,S. K. Kang,E. J. Kim,H. Seo,S. Y. Kim,W. Kim,M. Niiyama,S. J. M. Peeters,H. Ray,C. Rott,I. Yu,H. Jeon,S. Jeon,D. E. Jung,S. B. Kim,E. Kwon,D. H. Lee,I. Stancu,M. Yeh
DOI: https://doi.org/10.1088/1748-0221/15/07/T07003
2020-05-26
Abstract:The JSNS$^{2}$ (J-PARC Sterile Neutrino Search at J-PARC Spallation Neutron Source) experiment aims to search for neutrino oscillations over a 24\,m short baseline at J-PARC. The JSNS$^{2}$ inner detector is filled with 17 tons of gadolinium-loaded liquid scintillator (LS) and both the intermediate $\gamma$-catcher and the optically separated outer veto are filled with un-loaded LS. Optical photons from scintillation are observed by 120 Photomultiplier Tubes (PMTs). A total of 130 PMTs for the JSNS2 experiment were both donated by other experiments and purchased from Hamamatsu. Donated PMTs were purchased around 10 years ago, therefore JSNS$^{2}$ did pre-calibration of the PMTs including the purchased PMTs. 123 PMTs demonstrated acceptable performance for the JSNS$^{2}$ experiment, and 120 PMTs were installed in the detector.
Instrumentation and Detectors