An Ex Vivo Human Kidney Perfusion Model Confirms Larger Microwave Ablation Dimensions After Transarterial Embolization

Carlos B. Ortiz,Kade Derrick,Annie Dang,Marina Borrego,Seiji Yamaguchi,Daniel Grosser,Leon Bunegin,John Walker,Jorge Lopera
DOI: https://doi.org/10.1016/j.jvir.2024.06.012
IF: 3.682
2024-06-21
Journal of Vascular and Interventional Radiology
Abstract:Purpose To utilize a novel ex vivo perfused human renal model and quantify microwave ablation (MWA) size differences in renal tissue when combining MWA with transarterial embolization (TAE). Materials and methods Human kidneys (n = 5) declined for transplantation were obtained and connected to a fluoroscopic-compatible ex vivo perfusion system. Two ablations—1 standard MWA, 1 TAE-MWA—were performed in each kidney for 2 minutes at 100 Watts using a MWA system (Solero Angiodynamics). MWA alone was performed in the upper pole. In the lower pole, MWA was performed after TAE with M0 LUMI microspheres (Boston Scientific) to achieve angiographic stasis. Ablation zones of coagulative necrosis were sectioned along the long axis and segmented for maximal short axis diameter (SAD) and long axis diameter (LAD) measurements. Results A total of 10 ablations (5 MWA, 5 TAE-MWA) were performed in five human kidneys. TAE-MWA resulted in significantly increased SAD, LAD, volume, and sphericity compared to standard MWA + SD with mean measurements as follows (5 standard MWA + SD vs 5 TAE-MWA, two-tailed t-test): SAD, 1.8 ± 0.1 cm vs 2.5 ± 0.1 cm (p < 0.001); LAD, 2.9 ± 0.3 cm vs 3.2 ± 0.1 cm (p = 0.039); volume, 5.0 ± 0.5 mL vs 11.0 ± 0.7 mL (p < 0.001); sphericity, 0.4 ± 0.2 vs 0.6 ± 0.1 (p = 0.049). Histology demonstrated no differences in TAE-MWA other than concentrated microspheres. Conclusion This study utilized a novel ex vivo human kidney perfusion model to confirm combined MWA-TAE significantly increases ablation size and spherical shape.
radiology, nuclear medicine & medical imaging,peripheral vascular disease
What problem does this paper attempt to address?