A 2D antiscatter grid and scatter sampling based CBCT method for online dose calculations during CBCT guided radiation therapy of pelvis

Farhang Bayat,Brian Miller,Yeonok Park,Zhelin Yu,Timur Alexeev,David Thomas,Kelly Stuhr,Brian Kavanagh,Moyed Miften,Cem Altunbas
2023-10-11
Abstract:Online dose calculations before radiation treatment have applications in dose delivery verification, plan adaptation, and treatment planning. We propose a novel CBCT imaging pipeline to enhance accuracy. Our approach aims to improve HU accuracy in CBCT images for more precise dose calculations. A quantitative CBCT pipeline was implemented, combining data correction strategies and scatter rejection, achieving high CT number accuracy. We evaluated the pipeline's effect using pelvis anatomy phantoms and found that dosimetric errors in quantitative CBCT-based dose calculations were minimal. In contrast, clinical CBCT and high-performance ASG CBCT-based plans showed significant errors. The proposed quantitative CBCT pipeline offers comparable dose calculation accuracy to the gold-standard planning CT, eliminating the need for density overrides and enabling precise dose delivery monitoring or online plan adaptations in radiation therapy.
Medical Physics
What problem does this paper attempt to address?