A compact cryogenic configurable slit unit for a multi-object infrared spectrograph:Design and Development of a prototype at TIFR

P. P. Madhwani,A. P. K. Kutty,B. Mookerjea,J. V. Parmar,V. N. Kurhade,S. L. D'Costa,P. Manoj,A. Surya
2023-09-01
Abstract:We present a cryogenic configurable slit unit (CSU) for a multi object infrared spectrograph with an effective field of view of 9.1 arcmin x 9.1 arcmin that was completely conceived and designed in the laboratory at TIFR. Several components of the CSU including the controller for the commercially procured piezo-walkers, controlled loop position sensing mechanism using digital slide callipers and a cryogenic test facility for the assembled prototype were also developed in-house. The principle of the CSU involves division of the field of view of the spectrometer into contiguous and parallel spatial bands, each one associated with two opposite sliding metal bars that can be positioned to create a slit needed to make spectroscopic observations of one astronomical object. A three-slit prototype of the newly designed CSU was built and tested extensively at ambient and cryogenic temperatures. The performance of the CSU was found to be as per specifications.
Instrumentation and Methods for Astrophysics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to design and develop a compact cryogenic configurable slit unit (CSU) for the multi - object infrared spectrometer (MOIS). Specifically, the research team aims to create a device that can accurately configure multiple slits in a cryogenic environment to meet the astronomical observation requirements in the near - infrared band (0.8 - 2.5 µm). This CSU needs to be able to operate under cryogenic conditions to suppress background thermal radiation and be able to configure slits quickly and flexibly before observation. ### Main problems and challenges: 1. **Operation in a cryogenic environment**: Since the observation in the near - infrared band requires cooling the slits to 120 K or lower to reduce the influence of background thermal radiation, the CSU must work properly in a cryogenic environment. 2. **High - precision positioning**: To ensure the accuracy of spectral observations, the CSU needs to be able to precisely control the position and width of the slits. The width of each slit can range from fully closed (0 mm) to fully open (90 mm), and the absolute positioning accuracy is required to reach ±10 µm. 3. **Requirements for multi - object observation**: The CSU needs to be able to configure slits for multiple celestial bodies simultaneously within the same field of view, thereby improving the observation efficiency. For example, MOSFIRE can observe up to 46 targets within a 6.′12×6.′12 field of view, and EMIR can observe up to 53 targets within a 6.′64×4′ field of view. 4. **Modularity and extensibility**: The design needs to have modular characteristics so that it can be applied to larger - aperture telescopes (such as 10 - meter - class telescopes) in the future and support a larger number of slits. ### Solutions: - **Piezoelectric walker drive mechanism**: Use the PI NEXACT - 310 piezoelectric walker as a driving device, which can provide precise linear motion in a cryogenic environment and avoid the backlash error problem of traditional stepper motors. - **Closed - loop position sensing mechanism**: Achieve high - precision closed - loop position control through digital calipers and grid - capacitance linear - synchronous sensors to ensure that the positioning accuracy of each slit reaches ±10 µm. - **Mechanical design optimization**: The design of the slit bar adopts a special cross - sectional structure to ensure no light leakage in a cryogenic environment while maintaining smooth motion. - **Cryogenic test**: Design a special aluminum Dewar vessel in the laboratory for performance testing in a cryogenic environment to ensure the reliability of the CSU in actual observations. Through these design and technical means, the research team has successfully developed a three - slit prototype and conducted extensive tests at room temperature and in a cryogenic environment, verifying that its performance meets expectations. ### Future directions: The research team will continue to optimize the design, especially explore piezoelectric drivers suitable for lower temperatures, and evaluate whether a specially customized cryogenic piezoelectric driver is required to meet a larger movement range (100 mm). In addition, they will further improve the thermal background analysis to ensure the cryogenic performance of the entire system.