A Radiological Clip Design Using Ultrasound Identification to Improve Localization

Jenna Cario,Zhengchang Kou,Rita J. Miller,April Dickenson,Christine U. Lee,Michael L. Oelze
2024-02-02
Abstract:Objective: We demonstrate the use of ultrasound to receive an acoustic signal transmitted from a radiological clip designed from a custom circuit. This signal encodes an identification number and is localized and identified wirelessly by the ultrasound imaging system. Methods: We designed and constructed the test platform with a Teensy 4.0 microcontroller core to detect ultrasonic imaging pulses received by a transducer embedded in a phantom, which acted as the radiological clip. Ultrasound identification (USID) signals were generated and transmitted as a result. The phantom and clip were imaged using an ultrasonic array (Philips L7-4) connected to a Verasonics Vantage 128 system operating in pulse inversion (PI) mode. Cross-correlations were performed to localize and identify the code sequences in the PI images. Results: USID signals were detected and visualized on B-mode images of the phantoms with up to sub-millimeter localization accuracy. The average detection rate across 4,800 frames of ultrasound data was 93.0%. Tested ID values exhibited differences in detection rates. Conclusion: The USID clip produced identifiable, distinguishable, and localizable signals when imaged. Significance: Radiological clips are used to mark breast cancer being treated by neoadjuvant chemotherapy (NAC) via implant in or near treated lesions. As NAC progresses, available marking clips can lose visibility in ultrasound, the imaging modality of choice for monitoring NAC-treated lesions. By transmitting an active signal, more accurate and reliable ultrasound localization of these clips could be achieved and multiple clips with different ID values could be imaged in the same field of view.
Signal Processing
What problem does this paper attempt to address?
### Problems the Paper Aims to Solve The paper aims to address the issue of radiological clips used to mark breast cancer lesions losing visibility in ultrasound imaging during neoadjuvant chemotherapy (NAC). Specifically, the current marking clips may become difficult to identify via ultrasound during NAC treatment, which limits the success rate of preoperative radioactive seed localization guided by ultrasound. To solve this problem, the paper proposes a new design for radiological clips that can transmit encoded identification signals via ultrasound, enabling more accurate and reliable ultrasound localization. Additionally, multiple clips with different ID values can be identified within the same field of view. ### Key Points of the Solution 1. **Ultrasound Identification (USID) Signal**: - A clip design that can transmit an ultrasound signal encoded with an identification number. - The signal is wirelessly received and identified by the ultrasound imaging system. 2. **Testing Platform**: - A Teensy 4.0 microcontroller core is used to detect ultrasound imaging pulses and generate and transmit the USID signal. - Ultrasound transducers are embedded in a tissue-mimicking phantom to simulate radiological clips. 3. **Imaging System**: - A Philips L7-4 ultrasound array connected to a Verasonics™ Vantage 128 system is used for imaging in pulse inversion (PI) mode. - Cross-correlation processing is used to locate and identify the code sequences. 4. **Experimental Results**: - The USID signal was detected and visualized in B-mode images, with localization accuracy reaching sub-millimeter levels. - The average detection rate was 93.0%. 5. **Significance**: - By transmitting active signals, more accurate and reliable ultrasound localization can be achieved. - Multiple clips with different ID values can be imaged within the same field of view, improving localization accuracy during surgery. ### Conclusion This new USID clip design generates identifiable, distinguishable, and locatable signals during imaging, significantly improving the visibility and reliability of radiological clips in ultrasound imaging. This is of great importance for monitoring lesion markers during NAC treatment.