A pulsed optically pumped Rb clock with a frequency stability below 10-15

Michele Gozzelino,Salvatore Micalizio,Claudio E. Calosso,Jacopo Belfi,Adalberto Sapia,Marina Gioia,Filippo Levi
DOI: https://doi.org/10.1038/s41598-023-39942-5
2023-08-29
Abstract:We present the frequency stability performances of a vapor cell Rb clock based on the pulsed optically pumping (POP) technique. The clock has been developed in the frame of a collaboration between INRIM and Leonardo SpA, aiming to realize a space-qualified POP frequency standard. The results here reported were obtained with an engineered physics package, specifically designed for space applications, joint to laboratory-grade optics and electronics. The measured frequency stability expressed in terms of Allan deviation is $1.2 \times 10^{-13}$ at 1 s and achieves the value of $6 \times 10^{-16}$ for integration times of 40000 s (drift removed). This is, to our knowledge, a record result for a vapor-cell frequency standard. In the paper, we show that in order to get this result, a careful stabilization of microwave and laser pulses is required.
Atomic Physics,Instrumentation and Detectors,Optics
What problem does this paper attempt to address?
This paper aims to solve the problem of improving the frequency stability of rubidium atomic clocks. Specifically, the research team has developed a rubidium atomic clock based on the Pulsed Optically Pumped (POP) technology, aiming to achieve better frequency stability than existing technologies, especially the stability under long - time integration. The paper reports that the frequency stability of this rubidium atomic clock reaches an Allan deviation of \(1.2\times 10^{- 13}\tau^{-1/2}\), and reaches a level of \(6\times 10^{-16}\) at an integration time of 40,000 seconds (after removing the drift), which is the best - known result for rubidium vapor - cell frequency standards. To achieve this performance, the research particularly emphasizes the importance of precise control and stabilization of microwave and laser pulses. Through these technological improvements, the research team has successfully reduced the optical frequency shift effect, thereby significantly improving the long - term stability of the rubidium atomic clock. In addition, this research also shows how to improve short - term stability and signal - to - noise ratio by optimizing laser intensity, and how to reduce the influence of optical and microwave power broadening effects through the central Ramsey fringe in the Ramsey interference method. These technological advances are of great significance for the performance improvement of rubidium atomic clocks in future space applications.