Feasibility of dual-energy CBCT material decomposition in the human torso with 2D anti-scatter grids and grid-based scatter sampling

Cem Altunbas
2023-06-23
Abstract:Background: Dual-energy (DE) imaging techniques in cone-beam computed tomography (CBCT) have potential clinical applications, including material quantification and improved tissue visualization. However, the performance of DE CBCT is limited by the effects of scattered radiation, which restricts its use to small object imaging. Purpose: This study investigates the feasibility of DE CBCT material decomposition by reducing scatter with a 2D anti-scatter grid and a measurement-based scatter correction method. Methods: A 2D anti-scatter grid prototype was utilized with a residual scatter correction method in a linac-mounted CBCT system to investigate the effects of robust scatter suppression in DE CBCT. Scans were acquired at 90 and 140 kVp using phantoms that mimic head, thorax, and abdomen/pelvis anatomies. The effect of a 2D anti-scatter grid with and without residual scatter correction on iodine concentration quantification and contrast visualization in VME images was evaluated. Results: In CBCT images, a 2D grid with or without scatter correction can differentiate iodine and water after DE processing in human torso-sized phantom images. However, iodine quantification errors were up to 10 mg/ml in pelvis phantoms when only the 2D grid was used. Adding scatter correction to 2D-grid CBCT reduced iodine quantification errors below 1.5 mg/ml in pelvis phantoms, comparable to iodine quantification errors in multidetector CT. Conclusions: This study indicates that accurate DE decomposition is potentially feasible in DE CBCT of the human torso if robust scatter suppression is achieved with 2D anti-scatter grids and residual scatter correction. This approach can potentially enable better contrast visualization and tissue and contrast agent quantification in various CBCT applications.
Medical Physics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is the feasibility of dual - energy Cone - Beam Computed Tomography (DE CBCT) material decomposition in the human trunk area, especially when using a 2D anti - scatter grid and the Grid - based Scatter Sampling (GSS) method to reduce the influence of scatter. Specifically, the research focuses on the application of iodine quantification accuracy and Virtual Monoenergetic (VME) imaging in phantoms that mimic the head, chest, abdomen, and pelvis anatomical structures. ### Research Background Dual - energy imaging technology has potential clinical application value in Cone - Beam Computed Tomography, such as material quantification and improved tissue visualization. However, due to the influence of scattered radiation, the performance of DE CBCT is limited, which makes its application mainly limited to small - object imaging. ### Research Objectives This paper aims to explore the feasibility of DE CBCT material decomposition by using a 2D anti - scatter grid and a measurement - based scatter correction method to reduce scatter. The research specifically focuses on the performance of iodine quantification accuracy and VME imaging in phantoms that mimic different anatomical parts (head, chest, abdomen, pelvis). ### Methods - **Scatter Suppression**: Use a 2D anti - scatter grid prototype and combine it with the Residual Scatter Correction method (GSS) to conduct experiments in a CBCT system installed on a linear accelerator. - **Dual - Energy Processing**: Acquire CBCT images of phantoms at 90 kVp and 140 kVp and decompose them into iodine and water - based material images. - **Experimental Setup**: Use multiple phantoms, including phantoms that mimic the head, chest, abdomen, and pelvis anatomical structures, with different concentrations of iodine solution placed in each phantom. - **Data Analysis**: Evaluate the improvement effects of the 2D anti - scatter grid and GSS on iodine concentration quantification accuracy and VME imaging contrast visualization. ### Results - **Scatter Suppression Effect**: The 2D anti - scatter grid and GSS can significantly reduce scatter and improve the quantification accuracy of iodine concentration. In particular, in the pelvis phantom, when only using the 2D anti - scatter grid, the iodine quantification error is as high as 10 mg/ml; after combining with GSS, the error is reduced to less than 1.5 mg/ml, which is equivalent to the error of multi - detector CT (MDCT). - **Contrast Visualization**: Although no significant signal - to - noise ratio (CNR) improvement was observed in the VME CBCT images, in the 40 keV VME images, the contrast visualization was significantly better than that in the 90 kVp and 140 kVp CBCT images. ### Conclusions The research shows that if effective scatter suppression is achieved through the 2D anti - scatter grid and the Residual Scatter Correction method, DE CBCT material decomposition in the human trunk area is feasible. This method can potentially improve contrast visualization and the quantification accuracy of tissues and contrast agents, and thus play an important role in various CBCT applications.