Multi-energy CT imaging for large patients using dual-source photon-counting detector CT

Shengzhen Tao,Jeffrey F Marsh,Ashley Tao,Greg J Michalak,Kishore Rajendran,Cynthia H McCollough,Shuai Leng
DOI: https://doi.org/10.1088/1361-6560/ab99e4
IF: 3.5
2020-08-31
Physics in Medicine and Biology
Abstract:Multi-energy CT imaging of large patients with conventional dual-energy (DE)-CT using an energy-integrating-detector (EID) is challenging due to photon starvation-induced image artifacts, especially in lower tube potential (80–100 kV) images. Here, we performed phantom experiments to investigate the performance of DECT for morbidly obese patients, using an iodine and water material decomposition task as an example, on an emulated dual-source (DS)-photon-counting-detector (PCD)-CT, and compared its performance with a clinical DS-EID-CT. An abdominal CT phantom with iodine inserts of different concentrations was wrapped with tissue-equivalent gel layers to emulate a large patient (50 cm lateral size). The phantom was scanned on a research whole-body single-source (SS)-PCD-CT (140 kV tube potential), a DS-PCD-CT (100/Sn140 kV; Sn140 indicates 140 kV with Sn filter), and a clinical DS-EID-CT (100/Sn140 kV) with the same radiation dose. Phantom scans were repeated five times on each system. The DS-PCD-CT acquisition was emulated by scanning twice on the SS-PCD-CT using different tube potentials. The multi-energy CT images acquired on each system were then reconstructed, and iodine- and water-specific images were generated using material decomposition. The root-mean-square-error (RMSE) between true and measured iodine concentrations were calculated for each system and compared. The images acquired on the DS-EID-CT showed severe artifacts, including ringing, reduced uniformity, and photon starvation artifacts, especially for low-energy images. These were largely reduced in DS-PCD-CT images. The CT number difference that was measured using regions-of-interest across field-of-view were reduced from 20.3 ± 0.9 (DS-EID-CT) to 2.5 ± 0.4 HU on DS-PCD-CT, showing improved image uniformity using DS-PCD-CT. Iodine RMSE was reduced from 3.42 ± 0.03 mg ml<sup>−1</sup> (SS-PCD-CT) and 2.90 ± 0.03 mg ml<sup>−1</sup> (DS-EID-CT) to 2.39 ± 0.05 mg ml<sup>−1</sup> using DS-PCD-CT. DS-PCD-CT out-performed a clinical DS-EID-CT for iodine and water-based material decomposition on phantom emulating obese patients by reducing image artifacts and improving iodine quantification (RMSE reduced by 20%). With DS-PCD-CT, multi-energy CT can be performed on large patients that cannot be accommodated with current DECT.
engineering, biomedical,radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is the image artifact problems encountered during multi - energy CT imaging of obese patients using traditional dual - energy CT (DECT), especially the image artifacts caused by photon starvation in low - tube - voltage (such as 80 - 100 kV) images. These problems include ring artifacts, decreased uniformity, and photon - starvation artifacts, which seriously affect the accuracy of CT values and, in turn, the accuracy of material quantification (such as iodine concentration measurement). To overcome these problems, the researchers designed a multi - energy CT imaging method based on dual - source photon - counting detectors (DS - PCD - CT) and experimentally verified its performance in iodine and water material decomposition tasks under simulated obese - patient conditions. Compared with the traditional dual - source energy - integrating - type detectors (DS - EID - CT), DS - PCD - CT can significantly reduce image artifacts, improve image uniformity, and the accuracy of iodine quantification. Specifically, the paper solves the problems through the following points: 1. **Using dual - source photon - counting detectors**: Different from traditional energy - integrating - type detectors, photon - counting detectors can distinguish the energy of each incident photon, thereby achieving multi - energy imaging in a single scan. 2. **Optimizing energy - threshold settings**: By selecting appropriate energy thresholds, DS - PCD - CT can obtain a wider energy distribution, thereby improving the energy separation between multi - energy data sets. 3. **Experimental verification**: Through abdominal CT phantom experiments simulating obese patients, the performance of DS - PCD - CT in reducing image artifacts, improving image uniformity, and the accuracy of iodine quantification was evaluated. In summary, this paper aims to improve the quality and accuracy of multi - energy CT imaging for obese patients by introducing DS - PCD - CT technology, thereby expanding the scope of application of multi - energy CT in clinical applications.