Evaluation of Multi-Pixel Photon Counters in Energy Determination for PET Imaging
M. Niu,C. -M. Kao,J. Zhu,P. Xiao,A. Long,R. G. Wagner,G. Drake,C. -T. Chen,Q. Xie
DOI: https://doi.org/10.1088/1748-0221/7/04/t04001
2012-01-01
Journal of Instrumentation
Abstract:The multi-pixel photon counter (MPPC), also known as the silicon photo-multiplier (SiPM), is a novel, solid-state photodetector containing an array of Geiger-mode photodiodes (referred to as microcells below). MPPC is attractive for positron emission tomography (PET) imaging due to its PMT-like high gain, fast timing response, good photon detection efficiency (PDE), insensitivity to magnetic fields, and reasonable dark-count rates at room temperature. The number of microcells that an MPPC contains is a design parameter that concerns the tradeoffs between the PDE, dynamic range, response time and other properties. Generally, for a given pixel size, a smaller number of microcells yields a higher PDE, which is favorable, but a smaller dynamic range, which is not favorable. In this paper, we study and compare the gains, pulse shapes, and energy resolutions of three 1 × 1 mm2 MPPCs that contain 1600, 400 or 100 microcells when coupled to a common PET scintillator LYSO. We employ two radiation sources with different energies to investigate the nonlinearity in the energy response and develop an energy calibration method. Our results show that although the 100-microcell MPPC has the best apparent energy resolution, the 400-microcell MPPC has the best calibrated energy resolution.