Abstract:Electrical impedance tomography (EIT) is a noninvasive medical imaging modality utilizing the current-density/voltage data measured on the surface of the subject. Calderón's method is a relatively recent EIT imaging algorithm that is non-iterative, fast, and capable of reconstructing complex-valued electric impedances. However, due to the regularization via low-pass filtering and linearization, the reconstructed images suffer from severe blurring and under-estimation of the exact conductivity values. In this work, we develop an enhanced version of Calderón's method, using {deep} convolution neural networks (i.e., U-net) {as an effective targeted post-processing step, and term the resulting method by deep Calderón's method.} Specifically, we learn a U-net to postprocess the EIT images generated by Calderón's method so as to have better resolutions and more accurate estimates of conductivity values. We simulate chest configurations with which we generate the current-density/voltage boundary measurements and the corresponding reconstructed images by Calderón's method. With the paired training data, we learn the deep neural network and evaluate its performance on real tank measurement data. The experimental results indicate that the proposed approach indeed provides a fast and direct (complex-valued) impedance tomography imaging technique, and substantially improves the capability of the standard Calderón's method.
Numerical Analysis,Machine Learning
What problem does this paper attempt to address?
The problem that this paper attempts to solve is: **How to improve the image reconstruction quality of the Calderón method in electrical impedance tomography (EIT)**. Specifically, the Calderón method is a fast, non - iterative algorithm for EIT, which can reconstruct complex - valued conductivity distributions. However, due to low - pass filtering and linearization processing, the images reconstructed by the Calderón method have serious blurring and underestimated conductivity values. To solve these problems, the authors propose an enhanced version of the Calderón method, namely the **Deep Calderón Method**, which uses a deep convolutional neural network (such as U - net) to post - process the images generated by the Calderón method in order to improve the resolution and the accuracy of conductivity estimation. ### Main problems and solutions 1. **Problems**: - The images reconstructed by the Calderón method have serious blurring. - The conductivity values are underestimated. 2. **Solutions**: - Use a deep convolutional neural network (especially U - net) as a post - processing step to improve the images generated by the Calderón method. - Train a U - net model so that it can learn from the low - quality images generated by the Calderón method and generate high - quality images. - Generate paired training data sets through simulated data and evaluate the model performance on real measurement data. ### Method overview - **Calderón method**: First, use the Calderón method to generate the initial conductivity distribution images. - **U - net post - processing**: Then use U - net to post - process these initial images to improve the image quality and the accuracy of conductivity estimation. - **Training data generation**: Generate paired training data sets by simulating current density/voltage boundary measurement data of different thoracic cavity configurations. - **Experimental verification**: Verify the effectiveness of the proposed method on simulated data and real water tank measurement data. ### Experimental results The experimental results show that the Deep Calderón method can significantly improve the image quality and the accuracy of conductivity estimation while maintaining fast calculation. Especially in the case of complex - valued targets, this method can not only sharpen the organ boundaries, but also detect the size and position of organs more accurately, and can even detect very small abnormal areas in the lungs. In conclusion, this paper aims to improve the image reconstruction effect of the Calderón method in EIT by combining deep learning techniques, thereby providing a faster, more direct and more accurate complex - valued impedance tomography technique.