Optimization of patient-specific range modulators for conformal FLASH proton therapy

Sylvain Deffet,Kevin Souris,Edmond Sterpin
2023-03-15
Abstract:Purpose: A promising approach to enable FLASH conformal proton therapy is to passively degrade a single energy layer using a patient-specific range modulator. We propose an innovative method to directly optimize the geometrical characteristics of the range modulator and the treatment plan with respect to user defined constraints, similarly to state-of-the-art IMPT inverse planning. Methods: The kind of range modulators proposed in this study is a voxelized object which can be placed in the CT for dose computation, which simplifies the simulation pipeline. Both the geometrical characteristics of the range modulator and the weights of the PBS spots were directly optimized with respect to constraints on the dose using a first-order method. A modified Monte Carlo dose engine was used to provide an estimate of the gradient of the relaxed constraints with respect to the elevation values of the range modulator. Results: Assessed on a head and neck case, dose conformity logically appeared to be significantly degraded compared to IMPT. We then demonstrated that this degradation came mainly from the use of a large range shifter and therefore from physical limitations inherent in the passive degradation of beam energy. The geometry of the range modulator, on the other hand, was shown to be very close to being optimal. PBS dose rates were computed and discussed with respect to FLASH objectives. Conclusions: The voxelized range modulators optimized with the proposed method were proven to be optimal on a head and neck case characterized by two rather large volumes, with irregular contours and variable depths. The optimized geometry differed from conventional ridge filters as it was arbitrarily set by the optimizer. This kind of range modulators can be directly added in the CT for dose computation and is well suited for 3D printing.
Medical Physics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is how to achieve intensity - modulated proton therapy (IMPT) that meets the FLASH dose rate by optimizing patient - specific range modulators in proton therapy. The current Pencil Beam Scanning (PBS) system is unable to provide IMPT treatment at the FLASH dose rate. The main reason is that it takes time to change the beam energy, resulting in an overly long treatment time and an inability to meet the requirements of a high - dose rate. The paper proposes an innovative method that directly optimizes the geometric characteristics of the range modulator and the treatment plan to make it meet user - defined constraints, similar to the state - of - the - art IMPT inverse - planning method. This method not only simplifies the simulation process but also can generate range modulators of any geometric structure that highly match the shape of the target volume and are suitable for 3D printing, thereby improving the accuracy and efficiency of treatment.