Signal Preemphasis Technique for 10–60 MHz High-Speed Implant Communication System With Bandwidth-Limited Small Antennas

Lijia Liu,Hiroaki Takagi,Kota Miyazaki,Jingjing Shi,Jianqing Wang
DOI: https://doi.org/10.1109/tap.2024.3380748
IF: 5.7
2024-05-03
IEEE Transactions on Antennas and Propagation
Abstract:Real-time and high-quality control of implant robots requires wireless communication with tens of megabits per second high data rate and communication distance deeper than 10 cm within the human body. However, the frequency-dependent human tissue properties and the small implant antenna's narrow bandwidth can lead to severe channel distortion and bandwidth limitations, restricting the transmission speed and distance significantly. This study aims to achieve a high data rate of 20 Mb/s for an implant communication system equipped with narrow-band (NB) small antennas, within the 10–60 MHz human body communication (HBC) band. A signal preemphasis technique is proposed to mitigate channel distortion and overcome bandwidth limitations by precompensating the channel response while shaping the waveform of the transmitted (Tx) signal. By the proposed preemphasis technique, numerical simulation with an anatomical human body model shows that the receiving bandwidth is increased from 9.0 to 46.5 MHz, while reliable communication distance is achieved to 8 cm within the body model. Moreover, experimental measurement with a tissue-equivalent liquid phantom shows that the same antenna as in the simulation realizes a communication distance of 6 cm, while a better-designed wideband antenna enhances the distance to 10 cm.
telecommunications,engineering, electrical & electronic
What problem does this paper attempt to address?