Y Abe,T Chinzei,K Imachi,K Mabuchi,K Atsumi,I Fujimasa
Abstract:An artificial heart (AH) driving system, in which a sac or diaphragm type blood pump is liquid gas driven, is designed. The working mechanism of this system is as follows: 1) liquid gas is used for the driving source; 2) a liquid gas is stored in its liquid state in the circuit; 3) a liquid gas is vaporized, and the vaporizing pressure squeezes the blood pump, causing ejection of blood; 4) vaporized gas is aspirated and compressed by a small compressor to liquefaction through the heat exchanger, then negative pressure is applied to the blood pump and blood is aspirated; and 5) the blood pump is driven in this closed cycle. To demonstrate the mechanism of this system, a prototype was developed using Freon 114 as the liquid gas. In this system, the maximum flow of the AH at a 100 pulse per minute rate, was about 6.9 L/min, using a 90 ml sac type blood pump. The advantages of this AH driving mechanism are as follows: 1) a small system is available because pressure chambers are not necessary; 2) a biventricular system is available, with a single compressor; 3) no compliance chamber is necessary if the system is small enough to be implanted.