Single-exposure dual-energy computed radiography

B K Stewart,H K Huang
DOI: https://doi.org/10.1118/1.596479
Abstract:This paper focuses on analysis and development of a single-exposure dual-energy digital radiographic method using computed radiography (Fuji FCR-101 storage phosphor system). A detector sandwich consisting of storage phosphor imaging plates and an interdetector filter is used. The goal of this process is to provide a simple dual-energy method using typical plane-projection radiographic equipment and techniques. This approach exploits the transparency of the storage phosphor plates, using radiographic information that would be otherwise lost, to provide energy selective information essentially as a by-product of the radiographic examination. In order to effectively make use of the large dynamic range of the storage phosphor imaging plates (10,000:1), a computed radiography image reading mode of fixed analog-to-digital converter gain and variable photomultiplier sensitivity provides image data which can be related to relative incident exposure for export to the decomposition algorithm. Scatter rejection requirements necessitated crossed 12:1 grids for a field size of 36 x 36 cm. Optimal technique parameters obtained from computer simulation through minimization of the aluminum and Plexiglas equivalent image uncertainty under conditions of constant absorbed does resulted as: 100 kVp using a 0.15-mm-thick tin (Sn) interdetector filter for the lung field. This yields a surface exposure of 23 mR and a surface absorbed dose of 0.26 mGy for a 23-cm-thick chest. Clinical application in evaluation of the solitary pulmonary nodule is discussed, along with an image set demonstrating this application.
What problem does this paper attempt to address?