Analytical techniques for quantifying (15)N/(14)N of nitrate, nitrite, total dissolved nitrogen and ammonium in environmental samples using a gas chromatograph equipped with a quadrupole mass spectrometer

Kazuo Isobe,Yuichi Suwa,Junko Ikutani,Megumi Kuroiwa,Tomoko Makita,Yu Takebayashi,Muneoki Yoh,Shigeto Otsuka,Keishi Senoo,Masayuki Ohmori,Keisuke Koba
DOI: https://doi.org/10.1264/jsme2.me10159
Abstract:The measurement of (15)N concentrations in environmental samples requires sophisticated pretreatment devices and expensive isotope-ratio mass spectrometry (IRMS). This report describes the use of a gas chromatograph equipped with a quadrupole-type mass spectrometer (GC/MS) to measure (15)N concentrations of ammonium, nitrate, nitrite, and total dissolved nitrogen (TDN) in distilled water, a 2 M KCl solution and a 0.5 M K(2)SO(4) solution. The system measures nitrous oxide (N(2)O) that is ultimately converted from the target N compound, requiring no special apparatus such as a purge-and-trap pretreatment device. It uses a denitrifier lacking N(2)O reductase, which produces N(2)O from nitrate. Persulfate oxidation was applied to convert TDN to nitrate, while additional pretreatment with ammonia diffusion was required for ammonium prior to the persulfate oxidation. Up to 100 samples can be measured daily using the system. We can generally run (15)N measurements with only 1-10 mL of sample for each chemical species of N, a volume 1/10-1/100 times smaller than the amount necessary for conventional methods. Our method is useful for measuring (15)N with GC/MS, offering greater convenience than IRMS.