Proof-of-Concept Validation of Noninvasive Detection of Cortical Spreading Depolarization with High Resolution Direct Current-Electroencephalography

Benjamin R. Brown,Samuel J. Hund,Kirk A. Easley,Eric L. Singer,C. William Shuttleworth,Andrew P. Carlson,Stephen C. Jones
DOI: https://doi.org/10.1101/2024.11.12.24311133
2024-11-13
Abstract:Background/Objective: Cortical spreading depolarization (SD) is increasingly recognized as a major contributor to secondary brain injury. Monitoring SDs could be used to institute and guide SD-based therapeutics if noninvasive detection methods were available. Our primary objective is to use a high density array of electrodes to compare scalp direct current (DC)-shifts to SDs detected by gold standard electrocorticography (ECoG) to establish proof-of-concept validation that scalp DC-potentials can potentially provide noninvasive SD detection with a secondary objective of assessing usability and artifact tolerance. Methods: An 83x58 mm thermoplastic elastomer array with 29 embedded 6-mm diameter Ag/AgCl 1-cm spaced electrodes, the CerebroPatch™ Proof-of-Concept Prototype, was adhesively placed on the forehead with an intervening electrode gel interface to record DC-electroencephalography in normal volunteers and severe acute brain injury patients in the neuro-intensive care unit some with and some without invasive subdural ECoG electrodes. The scalp and ECoG voltages were collected by a Moberg® Advanced ICU Amplifier. Artifacts were visually identified and usability issues were recorded. SD was scored on ECoG using standard criteria of DC shift with associated suppression of high frequency activity with propagation across the electrode. A six parameter Gaussian plus quadratic baseline model was used to produce time-course ECoG and scalp electrode channel plots and heat-map movies of scalp voltages. The similarity of the noninvasive scalp and invasive ECoG DC-shift time-courses was compared via the Gaussian fit parameters and confirmed if the Coefficient-of-Determination exceeded 0.80. Results: Usability and artifact issues obscured most scalp Prototype device data except for 38 of the 140 ECoG-coded SDs over a period of 11 days in one sub-arachnoid hemorrhage patient. 26 of these DC-shifts were in readable, artifact free portions of scalp recordings and 24 had an acceptable, >0.80 Coefficient-of-Determination (0.98 [0.02], median [IQR]) between invasive ECoG and noninvasive Prototype device DC-shifts. These data suggest that these scalp DC-shifts (peak 457 ± 69 µV [mean ± StD], full-width-half maximum 70.9 ± 5.92 sec, area 18.7 ± 2.76 cm ) depicted in the heat-map movies represent noninvasively detected SDs. Conclusions: These results suggest that noninvasive SD detection is possible using scalp DC-potential signals with a high spatial resolution EEG array. Efforts to limit artifact and improve usability in DC-EEG detection are needed in order to improve the reliability of this approach and enable multi-modal monitoring for secondary brain injury.
What problem does this paper attempt to address?