Single particle cryo-electron microscopy with an enhanced 200 kV cryo-TEM configuration achieves near-atomic resolution

Lijia Jia,Eliza E. Ruben,Humberto J. Suarez,Shaun K. Olsen,Elizabeth Victorina Wasmuth
DOI: https://doi.org/10.1101/2024.05.07.593029
2024-05-10
Abstract:Single particle cryogenic electron microscopy (cryo-EM) as a structural biology methodology has become increasingly attractive and accessible to investigators in both academia and industry as this ever-advancing technology enables successful structural determination of a wide range of protein and nucleic acid targets. Although data for many high resolution cryo-EM structures are still obtained using a 300 kV cryogenic transmission electron microscope (cryo-TEM), a modern 200 kV cryo-TEM equipped with an advanced direct electron detector and energy filter is a cost-effective choice for most single particle applications, routinely achieving sub 3 angstrom resolution. Here, we systematically evaluate performance of one such high-end configuration, a 200 kV Glacios microscope coupled with a Falcon 4 direct electron detector and Selectris energy filter (Glacios-F4-S). First, we evaluated data quality on the standard benchmarking sample, rabbit muscle aldolase, using three of the most frequently used cryo-EM data collection software: SerialEM, Leginon and EPU, and found that, despite sample heterogeneity, all final reconstructions yield same overall resolutions of 2.6 angstrom and map quality when using either of the three software. Furthermore, comparison between Glacios-F4-S and a 300 kV cryo-TEM (Titan Krios with Falcon 4) revealed nominal resolution differences in overall reconstructions of a reconstituted human nucleosome core particle, achieving 2.8 and 2.5 angstrom, respectively. Finally, we performed comparative data analysis on the human RAD51 paralog complex, BCDX2, a four-protein complex of approximately 150 kilodaltons, and found that a small dataset of 1,000 micrographs was sufficient to generate a 3.3 angstrom reconstruction, with sufficient detail to resolve co-bound ligands, AMP-PNP and Mg+2. In summary, this study provides evidence that the Glacios-F4-S operates equally well with all standard data collection software, and is sufficient to obtain high resolution structural information of novel macromolecular complexes, readily acquiring single particle data rivaling that of 300 kV cryo-TEMs.
Biophysics
What problem does this paper attempt to address?
The main objective of this paper is to evaluate the performance of a high-end 200 kV cryo-transmission electron microscope (cryo-TEM) configuration (i.e., the Glacios microscope equipped with a Falcon 4 direct electron detector and a Selectris energy filter, abbreviated as Glacios-F4-S) in single-particle cryo-electron microscopy (cryo-EM) and compare it with a conventional 300 kV cryo-TEM. Specifically, the paper aims to address the following issues: 1. **Comparison of different data acquisition software**: Evaluate the data quality and resolution of Glacios-F4-S on the standard sample rabbit muscle aldolase using three commonly used data acquisition software (EPU, Leginon, and SerialEM). 2. **Performance comparison with 300 kV cryo-TEM**: Compare the overall resolution of Glacios-F4-S with the 300 kV Titan Krios in reconstructing the human nucleosome core particle. 3. **Structural analysis of small molecular complexes**: Demonstrate that Glacios-F4-S can obtain high-resolution structures of the human RAD51 homologous complex BCDX2 with a relatively small amount of data (≤1,000 micrographs) and clearly show the bound ligand AMP-PNP and Mg²⁺. Overall, this study aims to demonstrate that Glacios-F4-S, as a cost-effective solution, can rival 300 kV cryo-TEM in single-particle cryo-EM applications and obtain high-quality structural information.