Toward a Transportable Optical Clock with a Systematic Uncertainty of 4.8 × 10 −18

Mengyan Zeng,Yao Huang,Baolin Zhang,Yanmei Hao,Zixiao Ma,Ruming Hu,Huaqing Zhang,Zheng Chen,Miao Wang,Hua Guan,Kelin Gao
DOI: https://doi.org/10.1103/physrevapplied.19.064004
IF: 4.6
2023-06-02
Physical Review Applied
Abstract:We present the compact long-term nearly continuous operation of a room-temperature Ca+ -optical-clock setup towards a transportable clock, achieving an overall systematic uncertainty of 4.8 × 10−18 and an uptime rate of 97.8% over an 8-day period. The active liquid-cooling scheme is adopted, combined with precise temperature measurement with 13 temperature sensors both inside and outside the vacuum chamber to ensure the accurate evaluation of the thermal environment for the optical clock. The environmental temperature uncertainty is evaluated as 293.31(0.4) K, corresponding to a blackbody-radiation frequency-shift uncertainty of 4.6 × 10−18 , which is reduced more than twice compared to our previous work. Through frequency comparison of the room-temperature Ca+ optical clock with a cryogenic Ca+ optical clock, the overall uncertainty of the clock comparison is 7.5 × 10−18 , including a statistical uncertainty of 4.9 × 10−18 and a systematic uncertainty of 5.7 × 10−18 . This work provides a set of feasible implementations for high-precision transportable ion optical clocks. https://doi.org/10.1103/PhysRevApplied.19.064004 © 2023 American Physical Society
physics, applied
What problem does this paper attempt to address?