Abstract:Purpose Background suppression (BS) is recommended in arterial spin labeling (ASL) for improved SNR but is difficult to optimize in existing velocity‐selective ASL (VSASL) methods. Dual‐module VSASL (dm‐VSASL) enables delay‐insensitive, robust, and SNR‐efficient perfusion imaging, while allowing efficient BS, but its optimization has yet to be thoroughly investigated. Methods The inversion effects of the velocity‐selective labeling pulses, such as velocity‐selective inversion (VSI), can be used for BS, and were modeled for optimizing BS in dm‐VSASL. In vivo experiments using dual‐module VSI (dm‐VSI) were performed to compare two BS strategies: a conventional one with additional BS pulses and a new one without any BS pulse. Their BS performance, temporal noise, and temporal SNR were examined and compared, with pulsed and pseudo‐continuous ASL (PASL and PCASL) as the reference. Results The in vivo experiments validated the BS modeling. Strong positive linear correlations (r > 0.82, p
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?