Sensitivity enhanced NIR photoacoustic CO detection with SF6 promoting vibrational to translational relaxation process

Yingying Qiao,Liping Tang,Yang Gao,Fengtao Han,Chenguang Liu,Lei Li,Chongxin Shan
DOI: https://doi.org/10.1016/j.pacs.2022.100334
IF: 9.656
2022-03-01
Photoacoustics
Abstract:A challenge for slowly relaxing carbon monoxide (CO) molecules detection using photoacoustic spectroscopy (PAS) is to promote the vibration-translation (V-T) relaxation process. Addressing this challenge, a sensitivity enhanced photoacoustic CO sensor with sulfur hexafluoride (SF<sub>6</sub>) as the promotor is investigated and demonstrated. A 1568 nm near-infrared (NIR) laser diode and a customized optical amplifier are used as the excitation source to generate the photoacoustic signal. A differential photoacoustic cell is simulated and designed to obtain identical laminar flow distribution in the resonant cell to suppress the flow noise. The modulation frequency and added SF<sub>6</sub> volume ratio are optimized experimentally to achieve optimal sensitivity. Feasibility and performance of the CO sensor with a small amount of SF<sub>6</sub> as promotor is discussed and evaluated, obtaining a ~ 2 times improvement of signal value compared to the one with pure N<sub>2</sub> background and resulting in a minimum detection limit of 467.5 ppb for CO detection.
engineering, biomedical,radiology, nuclear medicine & medical imaging,instruments & instrumentation
What problem does this paper attempt to address?