Implementation of a dual library for the long-range (3 km) determination of sulfur hexafluoride using portable stand-off Fourier transform infrared (FTIR) spectroscopy

Hyung Joon Byun,Hyunwoo Nam,Jong-Seon Kim
DOI: https://doi.org/10.1080/10739149.2024.2417092
2024-10-18
Instrumentation Science & Technology
Abstract:A critical limitation in developing portable Fourier transform-infrared (FTIR) stand-off chemical detectors is the library optimization which reduces the number of false alarms and maintains the performance. Ideally, a library should be established through high-resolution (0.5 cm −1 ) FTIR spectroscopy in the laboratory. However, due to the weight requirements of a portable FTIR stand-off detector, its optical system and mirror-moving range have limitations. Consequently, most portable FTIR stand-off detectors have been developed with a low resolution of approximately 4 to 16 cm −1 . In this study, we developed a portable FTIR stand-off detector with a dual library of SF 6 . Moreover, we investigated its detection efficiency depending on the single library condition through realistic long-range (3 km) SF 6 spectra in a field test.
chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?