Development and Application of Capillary Electrophoresis-Electrospray Ionization Mass Spectrometry System Based on A Coaxial Sheath Flow Interface
Liu Zhu,Nai-Qian Zhang,Pan Fang,Jian-Zhang Pan,Catherine Wong,Qun Fang
DOI: https://doi.org/10.19756/j.issn.0253-3820.201408
IF: 1.193
2021-01-01
Chinese Journal of Analytical Chemistry
Abstract:A coaxial sheath-flow capillary electrophoresis-electrospray ionization mass spectrometry (CE-ESIMS) interface based on dual capillaries was developed for performing rapid, efficient and sensitive analysis. By designing different electrode systems, the online combination between homemade interface and different ESI- MS systems was achieved. The CE-ESI-MS system utilized the suction effect to accomplish hydrodynamic injection without external force, and realized the introduction of picoliter samples. The interface had many advantages such as simple to build, low cost, good durability, minimal dead volume and low dilution effect based on optimized parameters. The spray stability was evaluated by continuous injection with a relative standard deviation (RSD) of 3. 4%. Therefore, in the nine consecutive separations of reserpine, the RSD of retention time, full Width at Half Maximum and peak area were 1. 9%, 9. 2% and 8. 2%, respectively. The present system was successfully applied to separate eighteen standard amino acids within 9 min. Moreover, the baseline separation of five standard peptides mixture was achieved within 9 min with the number of theoretical plates of 7. 27 x 10(5) plates/m. The present system also performed good performance in the analysis of cytochrome C trypsin digest.