Investigations on physical and biological range uncertainties in Krakow proton beam therapy centre

Antoni Rucinski,Jakub Baran,Giuseppe Battistoni,Agnieszka Chrostowska,Marco Durante,Jan Gajewski,Magdalena Garbacz,Kamil Kisielewicz,Nils Krah,Vincenzo Patera,Monika Pawlik-Niedźwiecka,Ilaria Rinaldi,Bozena Rozwadowska-Bogusz,Emanuele Scifoni,Agata Skrzypek,Francesco Tommasino,Angelo Schiavi,Pawel Moskal
DOI: https://doi.org/10.5506/APhysPolB.51.9
2019-10-26
Abstract:Physical and biological range uncertainties limit the clinical potential of Proton Beam Therapy (PBT). In this proceedings, we report on two research projects, which we are conducting in parallel and which both tackle the problem of range uncertainties. One aims at developing software tools and the other at developing detector instrumentation. Regarding the first, we report on our development and pre-clinical application of a GPU-accelerated Monte Carlo (MC) simulation toolkit Fred. Concerning the letter, we report on our investigations of plastic scintillator based PET detectors for particle therapy delivery monitoring. We study the feasibility of Jagiellonian-PET detector technology for proton beam therapy range monitoring by means of MC simulations of the $\beta^+$ activity induced in a phantom by proton beams and present preliminary results of PET image reconstruction. Using a GPU-accelerated Monte Carlo simulation toolkit Fred and plastic scintillator based PET detectors we aim to improve patient treatment quality with protons.
Medical Physics
What problem does this paper attempt to address?