Effect of the Current Intensity and Inter-Electrode Distance in Surface Electrical Stimulation: A FEM Simulation Study

Rodrigo Osorio,Francisco Saavedra,B. Andrews,P. Aqueveque
DOI: https://doi.org/10.1109/EMBC48229.2022.9871286
2022-07-01
Abstract:In functional electric stimulation, variables such as electrode size, shape, and inter-electrodes distance can produce different neural and functional responses. In this work, a computational model combining FEM and MRG axon models is implemented to replicate two experimental studies that compare the effect of changing inter-electrode distance when applying FES to induce knee flexion. One work affirms that the stronger torque happens for greater distances, while the other obtain its maximum at lower distances. Using a simplified computational model gave another study perspective to understand why these two stimulation methodologies obtain different results. According to our results, an anodic stimulation occurs with greater current intensities and inter-electrode distances. This anodic effect can activate other nerve or motor points in the vicinity of the anode, explaining that more muscle fibers are recruited and generate an increased torque. Clinical Relevance - This work gives another view to understanding how the distance between electrodes affects neural activation, which has implications for optimizing clinical and exercise protocols using electrical stimulation techniques.
Medicine,Computer Science,Engineering
What problem does this paper attempt to address?