Generating Synthesized Computed Tomography (CT) from Cone-Beam Computed Tomography (CBCT) using CycleGAN for Adaptive Radiation Therapy

Xiao Liang,Liyuan Chen,Dan Nguyen,Zhiguo Zhou,Xuejun Gu,Ming Yang,Jing Wang,Steve Jiang
DOI: https://doi.org/10.1088/1361-6560/ab22f9
2018-12-12
Abstract:Cone beam computed tomography (CBCT) images can be used for dose calculation in adaptive radiation therapy (ART). The main challenges are the large artefacts and inaccurate Hounsfield unit (HU) values. Currently, deformed planning CT images are often used for this purpose, although anatomical accuracy might be a concern. Ideally, we would like to convert CBCT images to CT images with artifacts removed or greatly reduced and HU values corrected while keeping the anatomical accuracy. Recently, deep learning has achieved great success in image-to-image translation tasks. It is very difficult to acquire paired CT and CBCT images with exactly matching anatomy for supervised training. To overcome this limitation, we developed and tested a cycle generative adversarial network (CycleGAN) which is an unsupervised learning method and does not require paired training datasets to synthesize CT images from CBCT images. The synthesized CT (sCT) images have been compared with the deformed planning CT (dpCT) showing visual and quantitative similarity with artifacts being removed and HU value errors being reduced from 71.78 HU to 27.98 HU. Dose calculation accuracy using sCT images has been improved over the original CBCT images, with the average Gamma Index passing rate increased from 95.4% to 97.4% for 1 mm/1% criteria. A deformable phantom study has been conducted and demonstrated better anatomical accuracy for sCT over dpCT.
Medical Physics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is the main challenges encountered in dose calculation using Cone Beam Computed Tomography (CBCT) images in Adaptive Radiation Therapy (ART), namely, a large number of artifacts in CBCT images and inaccurate Hounsfield Unit (HU) values. These problems lead to inaccurate dose calculations based on CBCT images and affect the treatment effect. Currently, deformed planning CT (dpCT) images are often used for this purpose, but the accuracy of anatomical structures may be a problem. Ideally, it is hoped to convert CBCT images into CT images while removing or greatly reducing artifacts and correcting HU values while maintaining the accuracy of anatomical structures. To solve this problem, the author developed and tested a Cycle Generative Adversarial Network (CycleGAN), an unsupervised learning method that can synthesize CT images from CBCT images without paired training data sets. Through this method, the author aims to improve the accuracy of dose calculation, especially when using synthesized CT (sCT) images. Compared with the original CBCT images, the average Gamma index pass rate has increased from 95.4% to 97.4% (1 mm/1% standard). In addition, through the study of deformable phantoms, it has been proved that sCT is superior to dpCT in terms of anatomical structure accuracy.