A Monte-Carlo-based and GPU-accelerated 4D-dose calculator for a pencil-beam scanning proton therapy system

Mark D. Pepin,Erik Tryggestad,H. Wan Chan Tseung,Jedediah E. Johnson,Michael G. Herman,Chris Beltran
DOI: https://doi.org/10.1002/mp.13182
2018-09-13
Abstract:Purpose: The presence of respiratory motion during radiation treatment leads to degradation of the expected dose distribution, both for target coverage and healthy-tissue sparing, particularly for techniques like pencil-beam scanning proton therapy which have dynamic delivery systems. While tools exist to estimate this degraded four-dimensional (4D) dose, they typically have one or more deficiencies such as ... Methods: To quickly compute the 4D-dose, the three main tasks of the calculator were run on graphics processing units (GPUs). These tasks were: simulating the delivery of the plan using measured delivery parameters to distribute the plan amongst 4DCT phases characterizing the patient breathing, using an in-house Monte Carlo simulation (MC) dose calculator to determine the dose delivered to each breathing phase, and accumulating the doses from the various breathing phases onto a single phase for evaluation. The accumulation was performed by individually transferring the energy and mass of dose-grid subvoxels, a technique models the transfer of dose in a more physically realistic manner. The calculator was run ... Results: 4D doses were successfully computed for the three test cases with computation times ranging from 4-6 min on a server with eight NVIDIA Titan X graphics cards; the most time-consuming component was the MC dose engine. The subvoxel-based dose-accumulation technique produced stable 4D-dose distributions at subvoxel scales of 0.5-1.0 mm without impairing the total computation time. The uncertainties in the beam-delivery simulation ... Conclusions: A MC-based and GPU-accelerated 4D-dose calculator was developed to estimate the effects of respiratory motion on pencil-beam scanning proton therapy treatments. The calculator can currently be used ...
Medical Physics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is the degradation of dose distribution caused by respiratory motion during radiotherapy, especially when using proton pencil - beam scanning treatment technology. Specifically, existing tools have some flaws in estimating this four - dimensional (4D) dose, such as not including the specific effects of dynamic transport systems, using analytical dose calculation methods, and adopting non - physical dose accumulation methods. Therefore, this study aims to develop a clinically practical 4D dose calculator to address the above - mentioned deficiencies and accurately assess the impact of respiratory motion on proton pencil - beam scanning treatment. The method proposed in the paper is achieved through the following steps: 1. **Simulation plan transport**: Use the measured transport parameters to assign the plan to different 4DCT phases that describe patient respiration. 2. **Monte Carlo dose calculation**: Use an in - house developed Monte Carlo (MC) simulation dose calculator to determine the dose distribution in each respiratory phase. 3. **Dose accumulation**: Accumulate the doses of different respiratory phases onto a single phase for evaluation, and model dose delivery in a more physical way by separately transferring the energy and mass of dose grid sub - voxels. This study evaluated various uncertainties through three test cases (for lung, esophagus, and liver tumors respectively) and demonstrated the effectiveness and practicality of the developed 4D dose calculator.