Long-term assessment of PC-IOL optic apposition in surface modified IOL and its impact on Nd:YAG capsulotomy rate and visual quality

Manpreet Kaur,Jeewan S Titiyal,Jyoti Rawat
DOI: https://doi.org/10.1016/j.ajo.2024.09.029
IF: 5.488
2024-09-30
American Journal of Ophthalmology
Abstract:Purpose To assess apposition of posterior capsule (PC) to intraocular lens (IOL) optic in ultraviolet (UV)/ozone surface modified IOL and its impact on neodymium-doped yttrium aluminum garnet (Nd:YAG) laser capsulotomy rate and visual quality. Design Prospective Interventional Case Series Methods The study enrolled 100 eyes implanted with UV/ozone surface modified hydrophobic acrylic IOL during phacoemulsification. Primary outcome measure was PC-optic apposition on anterior segment optical coherence tomography (ASOCT). Secondary outcomes were Nd:YAG capsulotomy and visual quality (ray-tracing aberrometry). Follow up was performed on postoperative day (POD) 1, 1, 3, 6 months, 1 and 2 years. Results Complete PC-optic apposition was observed in 4% cases on POD1 which increased to 75% at 2 years. PC configuration in cases with PC-optic non-apposition was smooth (34.4%) or wavy (65.6%) on ASOCT; initial smooth PC was associated with significantly higher incidence of PC-optic apposition at 2 years (p=0.028). At 2 years, visual quality was significantly better in cases with complete PC-optic apposition in terms of Strehl ratio (p=0.029), modulation transfer function (MTF) (p=0.016), root mean square (RMS) aberrations (p=0.024) and higher order aberrations (HOAs) (p=0.043). Significant positive correlation was observed between height of interface fluid and total RMS [Pearson correlation coefficient 0.221; p=0.027] and HOAs [Pearson correlation coefficient 0.198; p=0.048].PCO developed in 7% (7/100); of these, 6 cases had persistent PC-optic non-apposition. Nd:YAG laser capsulotomy was required in 3% (3/100) cases; all these cases had persistent PC-optic non-apposition. Conclusion Long-term complete PC-optic apposition is observed in 75% cases implanted with UV/ozone surface modified IOLs. Complete PC-optic apposition is associated with lower Nd:YAG capsulotomy rate and superior visual quality.
ophthalmology
What problem does this paper attempt to address?