Source spatial averaging for the sensitivity measurement of multiple hydrophones in a tank

Qi Li,Jundong Sun,Dajing Shang,Rui Tang
DOI: https://doi.org/10.1016/j.apacoust.2020.107642
IF: 3.614
2021-01-01
Applied Acoustics
Abstract:<p>In this paper, we propose a novel calibration method for the batch measurement of hydrophone sensitivity in a tank. This calibration method is based on the source spatial averaging method of constructing a 'diffused sound field' in a small-volume tank. Source spatial averaging is implemented as follows: the transducer radiates steady-state signals whilst moving irregularly and slowly at a certain space, and average pressure is calculated equally at each position by average-processing the energy density in the reverberation control region. For demonstration, an experiment was conducted to calibrate six hydrophones in a glass tank with a dimension of 1.5 m × 0.9 m × 0.6 m and a frequency band range of 5–20 kHz. The measured readings were consistent with the factory-product identification. Error value was less than ± 0.5 dB, and standard uncertainty was less than 0.95 dB. Results demonstrate the accuracy and viability of the proposed batch calibration method based on source spatial averaging. The method can effectively and simultaneously calibrate multiple hydrophones in limited tank space and improve working efficiency.</p>