Investigation into the factors affecting accuracy of mass measurements on a time-of-flight mass spectrometer using Design of Experiment

Alice M-F Laures,Jean-Claude Wolff,Christine Eckers,Phil J Borman,Marion J Chatfield
DOI: https://doi.org/10.1002/rcm.2852
Abstract:The results of an investigation of the parameters which have the most significant effect on the accuracy of mass measurements on a quadrupole orthogonal acceleration time-of-flight mass spectrometer (q-oaToF) are reported. The influence of eight factors is investigated: ion abundances of reference and analyte compounds, mass difference between analyte and reference compounds, quality of calibration, number of reference acquisitions averaged and TDC (time-to-digital converter) settings (resolution, Np multiplier (number of pushes correction factor), minimum number of points, i.e. minimum acquisition width which defines a peak). To extract the maximum information from as few experiments as possible, a Design of Experiment approach was used. The data will be used as a basis for developing guidance on accurate mass measurement on q-oaToF instruments.
What problem does this paper attempt to address?