Modeling pressure distribution and heat in the body tissue and extract the relationship between them in order to improve treatment planning in HIFU

Saeed Reza Hajian,Ali Abbaspour Tehrani Fard,Majid Pouladian,Gholam Reza Hemmasi
DOI: https://doi.org/10.48550/arXiv.1610.08036
2016-10-26
Abstract:In high intensity focused ultrasound (HIFU) systems using non-ionizing methods in cancer treatment, if the device is applied to the body externally, the HIFU beam can damage nearby healthy tissues and burn skin due to lack of knowledge about the viscoelastic properties of patient tissue and failure to consider the physical properties of tissue in treatment planning. Addressing this problem by using various methods, such as MRI or ultrasound, elastography can effectively measure visco-elastic properties of tissue and fits within the pattern of stimulation and total treatment planning. In this paper, in a linear path of HIFU propagation, and by considering the smallest part of the path, including voxel with three mechanical elements of mass, spring and damper, which represents the properties of viscoelasticity of tissue, by creating waves of HIFU in the wire environment of MATLAB mechanics and stimulating these elements, pressure and heat transfer due to stimulation in the hypothetical voxel was obtained. Through the repeatability of these three-dimensional elements, tissue is created. The measurement was performed on three layers. The values of these elements for liver tissue and kidney of sheep in a practical example and outside the body are measured, and pressure and heat for three layers of liver and kidney tissue of an organism were obtained by applying ultrasound signals with a designed model. This action is repeated in three different directions, and the results are then compared with simulation software for ultrasound, as a reference to U.S. Food and Drug Administration (FDA) measures for HIFU, as well as comparisons of results with an operational method for an HIFU cell.
Medical Physics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is that in high - intensity focused ultrasound (HIFU) treatment, due to the lack of understanding of the viscoelastic properties of patient tissues and the failure to fully consider the physical properties of tissues in the treatment plan, the HIFU beam may damage nearby healthy tissues and burn the skin. To solve this problem, researchers have used methods such as MRI or ultrasonic elastography to effectively measure the viscoelastic properties of tissues and incorporate them into the stimulation mode and overall treatment plan. Specifically, the paper establishes a linear HIFU propagation path model, considering the smallest part in the path, that is, the voxel containing three mechanical elements: mass, spring and damper. These elements represent the viscoelastic properties of tissues. By creating HIFU waves in the MATLAB mechanical environment and stimulating these elements in this environment, the researchers obtained the pressure and heat transfer within the hypothetical voxel caused by stimulation. The tissue structure is reconstructed through the repeated construction of these three - dimensional elements. The experiment was carried out on three - layer tissues. Using actual examples of sheep liver and kidney tissues, the values of these elements were measured in vitro, and the pressure and heat values of the three layers of liver and kidney tissues were obtained by applying the designed model. This process was repeated in three different directions, and the results were compared with ultrasonic simulation software as the standard reference for HIFU by the US Food and Drug Administration (FDA), and also compared with the operation method of the HIFU unit. The main objective of the research is to improve the planning accuracy of HIFU treatment through accurate modeling and simulation, reduce damage to healthy tissues, and thus optimize the treatment effect.