High-extinction VIPA-based Brillouin spectroscopy of turbid biological media

Antonio Fiore,Jitao Zhang,Peng Shao,Seok Hyun Yun,Giuliano Scarcelli
DOI: https://doi.org/10.1063/1.4948353
2016-04-17
Abstract:Brillouin microscopy has recently emerged as powerful technique to characterize the mechanical properties of biological tissue, cell and biomaterials. However, the potential of Brillouin microscopy is currently limited to transparent samples, because Brillouin spectrometers do not have sufficient spectral extinction to reject the predominant non-Brillouin scattered light of turbid media. To overcome this issue, we developed a spectrometer composed of a two VIPA stages and a multi-pass Fabry-Perot interferometer. The Fabry-Perot etalon acts as an ultra-narrow band-pass filter for Brillouin light with high spectral extinction and low loss. We report background-free Brillouin spectra from Intralipid solutions and up to 100 microns deep within chicken muscle tissue.
Biological Physics,Optics,Quantitative Methods
What problem does this paper attempt to address?