The Dortmund Low Background Facility - Low-Background Gamma Ray Spectrometry with an Artificial Overburden

Holger Gastrich,Claus Gößling,Reiner Klingenberg,Kevin Kröninger,Till Neddermann,Christian Nitsch,Thomas Quante,Kai Zuber
DOI: https://doi.org/10.48550/arXiv.1512.01824
2015-12-07
Abstract:High-purity germanium (HPGe) detectors used for low-background gamma ray spectrometry are usually operated under either a fairly low overburden of the order of one meter of water equivalent (mw.e.) or a high overburden of the order of 100mw.e. or more, e.g. in specialized underground laboratories. The Dortmund Low Background Facility (DLB) combines the advantages of both approaches. The artificial overburden of 10mw.e. already shields the hadronic component of cosmic rays. The inner shielding, featuring a state-of-the-art neutron shielding and an active muon veto, enables low-background gamma ray spectrometry at an easy-accessible location at the campus of the Technische Universität Dortmund. The integral background count rate between 40keV and 2700keV is 2.528+-0.004counts/kg/minute. This enables activity measurements of primordial radionuclides in the range of some 10mBq/kg within a week of measurement time.
Instrumentation and Detectors,Nuclear Experiment
What problem does this paper attempt to address?