Abstract:Photoacoustic imaging can achieve high-resolution three-dimensional visualization of optical absorbers at penetration depths ~ 1 cm in biological tissues by detecting optically-induced high ultrasound frequencies. Tomographic acquisition with ultrasound linear arrays offers an easy implementation of single-side access, parallelized and high-frequency detection, but usually comes with an image quality impaired by the directionality of the detectors. Indeed, a simple translation of the array perpendicularly to its median imaging plane is often used, but results both in a poor resolution in the translation direction and in strong limited view artifacts. To improve the spatial resolution and the visibility of complex structures while keeping a planar detection geometry, we introduce, in this paper, a novel rotate-translate scanning scheme, and investigate the performance of a scanner implemented at 15 MHz center frequency. The developed system achieved a quasi-isotropic uniform 3D resolution of ~170 um over a cubic volume of side length 8.5 mm, i.e. an improvement in the resolution in the translation direction by almost one order of magnitude. Dual wavelength imaging was also demonstrated with ultrafast wavelength shifting. The validity of our approach was shown in vitro. We discuss the ability to enable in vivo imaging for preclinical and clinical studies.
Optics,Medical Physics
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to improve the resolution and field - of - view range in three - dimensional space for unilateral - access photoacoustic imaging (PAI), especially in high - frequency ultrasonic detection. The traditional linear array scanning method usually results in poor lateral resolution and severe field - of - view limitations. To solve these problems, the authors propose a new rotate - translate scanning scheme. By combining rotational and translational motions, the angular aperture of the ultrasonic linear array is expanded, thereby improving the spatial resolution of the image and the visibility of complex structures while keeping the planar detection geometry unchanged. Specifically, this research aims to: 1. **Achieve quasi - isotropic uniform 3D resolution**: By optimizing the scanning scheme, a resolution of approximately 170 micrometers is achieved within a cube volume (with a side length of 8.5 millimeters), which is nearly an order of magnitude higher than the resolution in the translation direction of the traditional method. 2. **Support multi - wavelength imaging**: Through ultrafast wavelength - switching technology, dual - wavelength imaging is achieved, which helps to reduce motion artifacts and improve image quality. 3. **Verify the feasibility of in - vivo imaging**: The effectiveness of this method is verified through in - vitro experiments, and its application potential in pre - clinical and clinical studies is discussed. In conclusion, this paper proposes a novel scanning scheme aiming to overcome the limitations of existing photoacoustic imaging techniques, especially in terms of unilateral access, high - resolution, and multi - wavelength imaging.