Normal Metal Hot-Electron Nanobolometer with Johnson Noise Thermometry Readout

Boris S. Karasik,Christopher B. McKitterick,Theodore J. Reck,Daniel E. Prober
DOI: https://doi.org/10.1109/TTHZ.2014.2370755
2014-11-05
Abstract:The sensitivity of a THz hot-electron nanobolometer (nano-HEB) made from a normal metal is analyzed. Johnson Noise Thermometry (JNT) is employed as a readout technique. In contrast to its superconducting TES counterpart, the normal-metal nano-HEB can operate at any cryogenic temperature depending on the required radiation background limited Noise Equivalent Power (NEP). It does not require bias lines; 100s of nano-HEBs can be read by a single low-noise X-band amplifier via a filter bank channelizer. The modeling predicts that even with the sensitivity penalty due to the amplifier noise, an NEP ~ 10$^{-20}$ - 10$^{-19}$ W/Hz$^{1/2}$ can be expected at 50-100 mK in 10-20 nm thin titanium (Ti) normal metal HEBs with niobium (Nb) contacts. This NEP is fairly constant over a range of readout frequencies ~ 10 GHz. Although materials with weaker electron-phonon coupling (bismuth, graphene) do not improve the minimum achievable NEP, they can be considered if a larger than 10 GHz readout bandwidth is required.
Instrumentation and Methods for Astrophysics,Mesoscale and Nanoscale Physics,Instrumentation and Detectors
What problem does this paper attempt to address?