A calibration procedure for beta-gamma coincidence detector-systems using four radioxenon spikes

Boxue Liu,Richard Britton,Seokryung Yoon,Ashley Vaughan Davies,Nikolaus Hermanspahn,Herbert Gohla,Jonathan Baré,Martin Kalinowski
DOI: https://doi.org/10.1016/j.apradiso.2024.111186
IF: 1.787
2024-01-12
Applied Radiation and Isotopes
Abstract:The determination of activity concentrations of the CTBT-relevant radioxenon relies on a robust calibration method. A procedure is outlined using four radioxenon spikes for beta-gamma detector-systems with 4π geometry. Detection efficiencies of beta-gamma coincidences in the net count calculation method, including the interference matrix between radioxenon and radon, are determined by three measurement channels: beta singles, gamma singles and beta-gamma coincidences, without reference activity values.