A simple configuration setup for compton suppression spectroscopy

N. X. Hai,P. D. Khang,N. N. Dien,V. H. Tan,N. D. Hoa
DOI: https://doi.org/10.48550/arXiv.1306.4110
2013-06-18
Abstract:The fast timing, standard timing and easy timing are popular timing configurations of compton suppression spectroscopy. Such spectroscopes always use a module of coincidence or time-to-amplitude converter (TAC). A compton suppression spectroscopy with semi-timing configuration is presented in this paper. The semi-timing configuration is relatively simple and easy system setup, especially this spectroscopy does not need to use module of coincidence or TAC. The performance of spectroscopy was tested and summarized. The count rate background, full peak efficiency and the ratios of area/background of peaks in suppressed and unsuppressed modes were comparative.