Direct leaf trajectory optimization for volumetric modulated arc therapy planning with sliding window delivery

Dávid Papp,Jan Unkelbach
DOI: https://doi.org/10.1118/1.4835435
2013-06-18
Abstract:We propose a novel optimization model for volumetric modulated arc therapy (VMAT) planning that directly optimizes deliverable leaf trajectories in the treatment plan optimization problem, and eliminates the need for a separate arc-sequencing step. In this model, a 360-degree arc is divided into a given number of arc segments in which the leaves move unidirectionally. This facilitates an algorithm that determines the optimal piecewise linear leaf trajectories for each arc segment, which are deliverable in a given treatment time. Multi-leaf collimator (MLC) constraints, including maximum leaf speed and interdigitation, are accounted for explicitly. The algorithm is customized to allow for VMAT delivery using constant gantry speed and dose rate, however, the algorithm generalizes to variable gantry speed if beneficial. We demonstrate the method for three different tumor sites: a head-and-neck case, a prostate case, and a paraspinal case. For that purpose, we first obtain a reference plan for intensity modulated radiotherapy (IMRT) using fluence map optimization and 20 equally spaced beam directions. Subsequently, VMAT plans are optimized by dividing the 360-degree arc into 20 corresponding arc segments. Assuming typical machine parameters (a dose rate of 600 MU/min, and a maximum leaf speed of 3 cm/sec), it is demonstrated that the quality of the optimized VMAT plans approaches the quality of the IMRT benchmark plan for delivery times between 3 and 4 minutes.
Medical Physics,Optimization and Control