A semi-packed gas chromatographic column with staggered elliptic cylindrical posts arrays

Boxin Chen,Yuchen Zhu,Shaojie Ma,Bin Zhao,Fei Feng
DOI: https://doi.org/10.1039/d3an02016a
2024-01-19
The Analyst
Abstract:The semi-packed gas chromatographic column has the advantages of high specific surface area and low column pressure. This paper reports that the stagnation regions formed in the adjacent posts along the channel of the semi-packed columns can decrease the area and height of chromatographic peaks, which makes it difficult to detect low-concentration mixed gases. A semi-packed column with staggered elliptic cylindrical posts arrays (SC-S) made by micro-electro-mechanical system technique was presented and compared the separation performance of SC-S with a semi-packed column with aligned elliptic cylindrical posts arrays (SC-A). The simulation results show that the width of stagnation regions in SC-S is 86.89% smaller than that in SC-A. The experiment results indicate that the area and height of chromatographic peaks increased as stagnation regions reduced. In the separations of the alkanes mixture from C8 through C10 with 10 ppm concentration, the chromatographic peak of decane is hardly to be identified in SC-A while the chromatographic peak in SC-S is still clearly visible. The chromatographic peak height of octane and nonane were increased by 65.06% and 130.00% in SC-S. The peak area of octane and nonane were increased by 120.45% and 168.18%.
chemistry, analytical
What problem does this paper attempt to address?