Cyclosporins monitoring by RIA and HPLC in liver transplantation: clinical correlation

Pierre E. Wallemacq,Michel Lesne,Jean‐Bernard Otte
DOI: https://doi.org/10.1111/j.1399-0012.1987.tb00704.x
1987-06-01
Clinical Transplantation
Abstract:Cyclosporine concentrations monitoring in hepatic transplantation is particularly difficult because the amount of cyclosporine metabolites increases during periods of impaired liver function. We have determined serum cyclosporine trough levels simultaneously and daily by radioimmunoassay (RIA) and liquid chromatography (HPLC) in 12 pediatric patients following liver transplantation. We have correlated the ratio of HPLC measurements of cyclosporine to RIA measurements (HPLC/RIA ratio) with the bilirubin levels during the 2 postoperative wk and the graft rejection periods. During periods of normal hepatic function we have observed HPLC/RIA ratio around 0.5. This ratio decreases significantly when serum bilirubin values exceed 50 mg/l. As long as the toxicity and the immunosuppressive activity of the cyclosporine metabolites are not better defined, cyclosporine monitoring should be performed by both RIA and HPLC methods during the 1st month posttransplantation.
surgery,transplantation
What problem does this paper attempt to address?